# Two PTEN antibody clones in breast cancer: high agreement in one centre, with clear limits

> In 150 hormone receptor-positive, HER2-negative breast cancers, two PTEN antibody clones gave the same loss or retained call in 149 cases. The data come from one centre, use adjudicated reads and one disagreement, so local verification is still needed.

Type: Briefing · Topic: Laboratory quality
Published: 2026-11-28T05:17:44+00:00
Updated: 2026-11-28T05:17:44+00:00
Canonical URL: https://www.pathologistsbrief.com/article/pten-antibody-concordance-breast-ihc
Publisher: Pathologist’s Brief
Source: https://doi.org/10.1186/s13000-026-01819-2
Author: Editor
Reviewer: Pathologist’s Brief editorial team
Review date: 2026-11-28

## Reporting

PTEN loss on immunohistochemistry is being considered as a predictive biomarker for pathway-directed treatment in advanced hormone receptor-positive, HER2-negative breast cancer, which is why laboratories are introducing the test. A practical question follows: when a laboratory uses a different antibody clone, will the loss or retained call stay the same?

Authors in China addressed this in a single-centre retrospective study. Shi et al. reported it in [Diagnostic Pathology, August 2026](https://doi.org/10.1186/s13000-026-01819-2).

The study included 150 invasive hormone receptor-positive, HER2-negative breast cancer surgical specimens collected from March 2024 to October 2025. All patients were female, aged 28 to 86 years, with a median age of 50 years. Of the tumours, 133 (88.7%) were invasive carcinoma of no special type.

Consecutive 4 micrometre sections were stained with each antibody, each on its own fully automated platform, following the manufacturers' protocols. Antibody and platform are therefore inseparable in this design. Normal mammary epithelium served as the internal positive control.

The read was binary, loss or retained. The scoring rule was taken from a published trial: loss meant that 90% or more of surviving tumour cells showed no specific cytoplasmic staining. Two senior pathologists read the slides independently and blinded. A third senior pathologist reviewed the cases on which they differed and reached a consensus.

The first antibody called loss in 35 cases (23.3%) and retained expression in 115 (76.7%). The second antibody called loss in 34 cases (22.7%) and retained expression in 116 (77.3%).

Taking the first antibody as the reference, the second antibody had a positive percent agreement of 97.1% (34/35), a negative percent agreement of 100% (115/115) and an overall percent agreement of 99.3% (149/150). The kappa statistic was 0.98 (95% CI 0.94 to 1.00). The cross-tabulation showed 34 cases of loss with both antibodies, 1 case of loss with the first antibody and retained with the second, 0 cases the other way round, and 115 retained with both.

These inter-antibody figures use the consensus reads after the third pathologist's adjudication.

Between the two readers, agreement for the first antibody was 146 of 150 cases (97.3%), with 4 discordant cases and a kappa statistic of 0.93 (95% CI 0.86 to 1.00). For the second antibody, the readers agreed on 148 of 150 cases (98.7%), with 2 discordant cases and a kappa statistic of 0.96 (95% CI 0.91 to 1.00). The third pathologist resolved all of these disagreements. The authors suggest that the high agreement between readers may relate to the binary scoring system, the clear scoring rule and the internal positive control.

Sequencing with a 56-gene targeted panel was done on the single case where the antibodies disagreed (loss with the first antibody, retained with the second). It found no pathogenic PTEN single nucleotide variants, insertions or deletions, or copy number variations. The authors add that immunohistochemistry and sequencing detect different biological levels. PTEN protein can be lost by non-genomic mechanisms such as epigenetic silencing, microRNA regulation or increased protein degradation, so the absence of genomic alterations does not explain the disagreement.

The authors list their own limitations: a small single-centre sample, only one discordant case so mechanisms of PTEN loss cannot be analysed, and no treatment follow-up so the link to clinical benefit is unknown. They call for multicentre and prospective validation.

## Interpretation

With 149 of 150 cases concordant, the result is encouraging for a laboratory that is validating a second clone. Several points temper it.

The inter-antibody figures use consensus reads after adjudication. Agreement before adjudication is not reported, so the contribution of adjudication cannot be judged.

Positive and negative agreement treat the first antibody as the reference. They measure agreement, not accuracy. The paper gives no confidence interval for positive agreement, which rests on only 35 loss cases.

Sequencing was done only on the single discordant case. The paper therefore has no sequencing-to-staining concordance, and the sequencing result neither supports nor refutes either antibody.

Antibody and platform were confounded. The paper cannot say whether the difference or the agreement comes from the clone or the platform, and it does not show that results are comparable across platforms.

For laboratories in India and elsewhere, the general lesson is practical. A laboratory introducing PTEN should verify on its own platform with its own cases. A fixed binary rule, an internal control, blinded double reading and a recorded process for resolving disagreements are sensible foundations. One centre's result does not replace that local verification.

## Opinion

This is a clearly framed, practical laboratory question, and the dataset is tidy and arithmetically consistent. The limitations paragraph is honest.

Still, one cut-off, one centre, binary reads by senior pathologists and one disagreement make "high agreement in one centre" the supportable claim. The abstract's conclusion about interchangeability runs ahead of the data, so this briefing does not repeat it.

## Limits

- Retrospective, single-centre study of 150 cases.
- Only one inter-antibody disagreement.
- Inter-antibody figures use consensus reads after adjudication.
- The first antibody was used as the reference.
- Antibody and platform were confounded.
- No sequencing-to-staining concordance, and no sequencing of concordant cases.
- Binary read with one scoring rule only.
- No clinical or treatment follow-up.
- No external or multicentre validation.

This briefing summarises one study. It is not a validated protocol and gives no advice on patient management.

## Frequently asked questions

### Did the two PTEN antibodies give the same result?
Almost always: they agreed in 149 of 150 cases after consensus reading. One case showed loss with the first antibody and retained expression with the second. The study was retrospective, from one centre, and did not report agreement before adjudication.

### Can a laboratory skip its own verification because of this study?
No, one centre's result does not replace local verification. Antibody and platform were inseparable here, so the study cannot show comparability across platforms. A laboratory should test on its own platform and cases, with a fixed rule, blinded double reading and a recorded way to resolve disagreements.

### What did sequencing of the discordant case show?
It found no pathogenic PTEN variants or copy number changes in that case. Only this one case was sequenced, so the paper has no sequencing-to-staining concordance. The authors note that PTEN protein can be lost by non-genomic mechanisms, so the result does not explain the disagreement.