Pathologist’s Brief
Reference · Digital pathology

Validating an AI tool in the pathology laboratory: the DPA recommendations, summarised

Quick answer

The Digital Pathology Association recommends validating the scanner and the AI separately, matching evidence to intended use, and qualified pathologist oversight. It separates technical performance, clinical meaning and improved patient outcomes.

Black-and-white photo of a slide scanner beside a rack of glass slides, a monitor with two matching tissue images and an agreement plot, and a clipboard log.
AI-generated illustration, Pathologist’s Brief
On this page
  1. Scope of the statement
  2. Three kinds of evidence
  3. Scanner first, then algorithm
  4. Accuracy, reliability and reference standards
  5. Quality control and minimum sample requirements
  6. Using AI within its intended use
  7. How strong are these recommendations?
  8. How to use this page
  9. Frequently asked questions

#Scope of the statement

Silberman et al., AI in Precision Oncology, first published 10 June 2026 present a Digital Pathology Association (DPA) recommendation statement covering scanning, image viewing and clinical AI interpretation.

Its scope is brightfield imaging of formalin-fixed, paraffin-embedded (FFPE) solid tissue. Cytology and immunofluorescence-based imaging are excluded.

#Three kinds of evidence

  • Analytical validation: does the system accurately and reliably measure its intended output under defined conditions, regardless of its clinical meaning?
  • Clinical validation: does the output accurately and reliably predict or correlate with a clinical condition, risk or outcome in the intended population and setting?
  • Clinical utility: can improved patient outcomes be attributed to using the test? Evidence may be direct or follow a validated chain of evidence linking the result, a treatment decision and a positive change in outcome.

The statement describes concordance with accepted diagnostic approaches and noninferiority studies as among the ways to build a chain of evidence. Technical accuracy alone does not establish clinical utility.

#Scanner first, then algorithm

R4: the statement recommends validating the scanning process separately from the AI algorithm. Once scanning is validated, it calls for validating the algorithm on images produced by that scanner. A scanner outside a regulatory-approved system should undergo laboratory validation in accordance with a consensus guideline.

R5: the statement calls for analytical and clinical validation for every scanner model on which the algorithm will be used clinically. Hardware, optics, resolution, colour calibration, compression and file format can affect image appearance and algorithm performance.

R6: the statement says interscanner concordance studies may support transfer from a validated scanner to another scanner, using the original scanner as the reference. Algorithm outputs from the same slides scanned on both systems are compared, using reference slides that reflect relevant sample variability.

The statement gives illustrative examples of Cohen's kappa >0.8 or categorical concordance of ≥95. These are examples, not universal acceptance thresholds. The metric and threshold depend on the task, with higher thresholds needed for higher-risk tasks.

R7: the statement calls for documenting adjustable scanner settings and conditions, including focus, compression and z-stacking. If a range is specified, it calls for validation across that range. Locked settings may be referenced as defaults while they remain unchanged.

#Accuracy, reliability and reference standards

R8: the statement says analytical validation should assess both accuracy and reliability. Accuracy concerns agreement with the appropriate reference standard. Reliability concerns consistency across repeated runs and varying conditions, such as staining batches, scanners or image artefacts.

R9: for predictions that pathologists can verify, the statement says cohort size should follow intended use. The general reference standard is a minimum of three pathologists, allowing majority voting when full consensus is absent.

The statement permits exceptions:

  • A single pathologist may suffice for a laboratory adding validation after manufacturer validation.
  • Research-use-only deployments may use a single pathologist, with the limitation acknowledged and fewer than three justified.
  • A tool customised to assist an individual pathologist may use that pathologist as the reference.

R10: for predictions that pathologists cannot visually verify, the statement calls for a reference standard corresponding to the predicted output. Examples include traditional molecular test results for molecular biomarker prediction and longitudinal patient outcomes for prognostic risk or therapeutic-response prediction. Virtual immunohistochemical staining may be compared with an adjacent immunohistochemical slide.

#Quality control and minimum sample requirements

R11: the statement recommends establishing quality systems across the entire testing process to prevent reporting invalid results caused by laboratory errors. It calls for identifying sources of variability, including batch effects, staining changes and scanner changes. Control methods and acceptance criteria should reflect intended use, specimen types and the reportable output range; intermittent control specimens or within-batch controls may be used.

R12: the statement calls for specifying minimum usable tissue area and cancer content for each algorithm. Small biopsies or sparse cancer may yield unreliable outputs. Tissue with artefacts that adversely affect validity, or whose effect is unknown, should not count as usable. The statement also calls for documenting which artefacts the algorithm can tolerate and its limitations.

#Using AI within its intended use

R3: the statement says all clinical AI requires qualified pathologist oversight and does not replace the pathologist. Even tools with defined inputs and outputs cannot remove the need to recognise rare diseases and unusual presentations.

R13: the statement recommends using the algorithm within its intended use, where it can improve interpretation or provide new information. A system intended as a second read should not become the effective primary read through reliance beyond that role.

The statement describes three categories of appropriate use:

  • Replacing or complementing an existing invasive, expensive or time-consuming test when sufficiently validated, addressed in R14.
  • Augmenting pathologist assessment of tasks whose outputs can be verified.
  • Predicting outcomes from features that pathologists cannot readily assess.

R15: the statement says assistance should be sufficiently validated and medically necessary. It calls for peer-reviewed evidence of improved diagnostic accuracy, reliability or throughput, analytical and clinical validation following its recommendations, and laboratory verification with appropriate quality control.

Its medical-necessity definition includes an anatomical pathology service ordered by the treating clinician and assessment by a board-eligible or board-certified pathologist or dermatopathologist. This terminology reflects the statement's regulatory and payer context.

#How strong are these recommendations?

The working group adapted GRADE, using High, Medium and Low evidence certainty and Strong or Conditional recommendation strength. Recommendations with Very Low confidence were excluded. Strong indicates confidence in improved patient outcomes; Conditional indicates moderate confidence that benefits outweigh costs.

Selected ratings are:

  • R5, scanner-specific validation: certainty High; strength Strong.
  • R7, scanner settings: certainty Medium; strength Strong.
  • R11, quality systems: certainty Medium; strength Strong.
  • R13, intended use: certainty Low; strength Strong.
  • R3, pathologist oversight: certainty N/A; strength Strong.

Evidence certainty and recommendation strength are therefore separate judgements. A Strong recommendation does not necessarily have High evidence certainty.

The search of publicly available sources was not a formal systematic review. The working group included pathologists and technologists from academia and industry.

#How to use this page

This page summarises a published recommendation statement written largely from a US regulatory and payer perspective. The full statement is best read alongside local regulations and accreditation requirements. Indian readers should check the requirements applicable to their own setting.

Frequently asked questions

Does a validated scanner make an AI algorithm validated?

No. The statement separates the scanning process from subsequent image analysis and calls for evidence for the algorithm on each scanner model used clinically. Interscanner concordance may support transfer, but settings, reference slides and task-dependent acceptance criteria still need to be addressed.

Must every reference set be reviewed by three pathologists?

No. The statement allows exceptions to its general three-pathologist reference standard. A single reviewer may suffice for additional local validation after manufacturer validation, a justified research-use-only deployment, or a tool customised for an individual pathologist. Predictions that cannot be visually verified need a reference corresponding to the output.

Does Strong mean the evidence certainty is High?

No. Recommendation strength and evidence certainty are separate judgements. The statement includes Strong recommendations with Low certainty, including intended use, and others with certainty recorded as N/A. Its adapted grading approach also excluded Very Low confidence recommendations, and the underlying search did not use a formal systematic review methodology.

Source

https://doi.org/10.1177/2993091X261455975 ↗

For education and professional reference. This does not give patient-specific advice or replace professional judgement, local protocols or consultation. Editorial standards.