# FDA proposal to reclassify oncology IHC tests is open for comment until 11 January 2027

> The US FDA proposes moving immunohistochemistry tests linked to approved oncology drugs from class III to class II with special controls. Comments are open until 11 January 2027.

Type: Briefing · Topic: Laboratory quality
Published: 2026-11-20T05:55:12+00:00
Updated: 2026-11-20T05:55:12+00:00
Canonical URL: https://www.pathologistsbrief.com/article/fda-ihc-companion-diagnostic-reclassification
Publisher: Pathologist’s Brief
Source: https://www.federalregister.gov/documents/2026/11/12/2026-23005/hematology-and-pathology-devices-reclassification-of-immunohistochemistry-test-systems-for-use-with
Author: Editor
Reviewer: Pathologist’s Brief editorial team
Review date: 2026-11-20

## Reporting

The FDA proposed order appeared in [the Federal Register on 12 November 2026](https://www.federalregister.gov/documents/2026/11/12/2026-23005/hematology-and-pathology-devices-reclassification-of-immunohistochemistry-test-systems-for-use-with). Under the proposed new section 21 CFR 864.1895, immunohistochemistry (IHC) test systems used with a corresponding approved oncology drug would move from class III (premarket approval) to class II with special controls. Premarket notification would still be required for new test systems.

The proposal covers 12 product codes. It includes companion diagnostics, which provide information essential to safe and effective use of the corresponding drug. It also includes tests cited in the drug labelling that give information about known benefits or risks without being essential to its use.

The FDA relied on data from 14 original premarket approval applications and reviewed adverse event reports and recalls. As of 15 July 2026, there were 48 reported events. Reports for over 70 percent of those events stated that there was no patient consequence or no known consequence. The recall review found no class I recalls, 10 class II recalls and 1 class III recall.

## Interpretation

Manufacturers would need to state how results are interpreted and reported, explain the choice of cut-offs or scoring criteria, and validate cut-offs using borderline samples around the clinical threshold. Analytical specificity studies would assess a range of normal and tumour tissues, interference and unintended binding to other targets, as applicable.

Reproducibility studies would use clinical specimens, including those near the threshold for classification, and assess variation across reagent lots, operators, instruments and days. They would test agreement between readers and consistency when the same reader scores a specimen again. Tests intended for use at more than one site would also need to show that results are consistent across sites.

Manufacturers would assess how much variation in test and specimen conditions the system can tolerate and how stable specimens remain, for example when stored as unstained cut sections.

Where appropriate, the labelling would specify that a qualified or appropriately trained reader should interpret results alongside other laboratory and pathology findings, relevant clinical information and proper controls. Companion diagnostic labelling would have to match the corresponding drug labelling.

Under the proposal, test systems already approved when a final order takes effect could continue to be marketed under their earlier approvals. The proposed effective date is 30 days after publication of any final order.

## Opinion

Testing reader agreement on borderline specimens speaks directly to a scoring problem pathologists face when a small difference can change the reported category. For laboratories in India and elsewhere that use these tests, the outcome is worth following for what it could mean for the performance information manufacturers provide.

## Limits

The proposal may change before any final order. It leaves specific cut-offs and numerical performance targets to individual test evaluations and retains premarket notification. Its controls apply to manufacturers of the covered test systems, with no direct requirements for laboratories or laboratory-developed tests.

The safety summary is based on reported events. The order gives no count of events that caused patient harm, so the reported totals do not support a conclusion about how often harm occurred.

## Frequently asked questions

### Does the proposal cover only companion diagnostics?
The proposal covers companion diagnostics and certain other oncology IHC tests. The distinction depends on the role of the test in the corresponding drug labelling. A companion diagnostic supplies information essential to safe and effective use, while another covered test can inform an assessment of known benefits or risks.

### What would a reproducibility study tell me about slide scoring?
A reproducibility study would show how consistently readers score clinical specimens. The proposed controls look at agreement between different readers and consistency on repeat readings by the same reader. Specimens close to the clinical threshold for classification must be included, where a small scoring difference can change the reported category.

### Would an already approved test need a new authorisation?
An already approved test could stay on the market under its existing approval if the proposal is finalised as written. That provision covers tests with approval in place when the final order takes effect. For new test systems, the proposed class II route would still include premarket notification.