# Whole-slide imaging for primary cytology reporting in a single-centre validation

> In a small validation dominated by non-gynaecological cytology, digital diagnoses agreed with glass in roughly 90% of comparisons when cases that were difficult even on glass were retained.

Type: Briefing · Topic: Digital pathology
Published: 2026-11-20T10:15:55+00:00
Updated: 2026-11-20T10:15:55+00:00
Canonical URL: https://www.pathologistsbrief.com/article/digital-cytology-primary-diagnosis-validation
Publisher: Pathologist’s Brief
Source: https://doi.org/10.1111/cyt.70063
Author: Editor
Reviewer: Pathologist’s Brief editorial team
Review date: 2026-11-20

## Reporting

Mistry et al., researchers in the United Kingdom, tested whole-slide imaging for primary cytology reporting in a small, single-centre validation. They selected 88 archived cases to resemble routine work, including 83 non-gynaecological and 5 cervical cases. The 201 slides included smears, liquid-based preparations, cytospins and a few cell blocks. The team scanned them at 40× equivalent magnification on a single scanner without z-stacking. [Cytopathology, March 2026](https://doi.org/10.1111/cyt.70063)

Two cytopathologists independently read each case digitally, without knowing the original glass diagnosis. Only one of the four readers had digital cytology reporting experience. Any mismatch with the original diagnosis triggered a glass re-review and a consensus meeting.

Both digital readings matched the original glass diagnosis in 65 of 88 cases. In the other 23, at least one reader differed from the original diagnosis. Glass review restored agreement in 8 cases and found a diagnostic discrepancy in 15, with six major and nine minor discrepancies.

After excluding those 15 cases, concordance was 95.1% (136/143, 95% confidence interval 89.8% to 97.8%), with 3 readings deferred to glass. With those cases included, concordance was 89.5% (154/172) when digital reads were compared with the glass re-review where available and the original diagnosis otherwise. Concordance was 89.8% (149/166) using the majority glass diagnosis as the reference. Both calculations excluded deferrals, and the majority analysis also omitted cases without a majority glass diagnosis.

Errors included missed carcinoma in three specimens, missed granulomatous inflammation and two false positives. Of the 4 deferrals to glass, 3 were smears. Thick areas were the main reason for deferral. Among the 5 cervical cases, 2 were fully concordant.

## Interpretation

The headline 95.1% depends on leaving out the 15 cases that were difficult even on glass. The roughly 90% concordance when those cases were retained is a fairer guide for laboratories planning validation.

The authors attributed most discordance to interpretation. On review of cases with agreement on glass, readers could see the diagnostic features in the digital images. Thick smears, three-dimensional cell clusters and sparse atypical cells were difficult to assess on screen.

All readers felt more confident with glass and reported that digital reporting took as long or longer. The cervical readers found screening on screen less practical because systematically covering the slide was cumbersome.

## Opinion

The study offers laboratory leads a template for local validation, particularly in how it records deferrals and reviews difficult cases. A laboratory should test each preparation type it uses, plan how readers can defer to glass or request a rescan, and keep deferred reads visible when reporting validation results. In the study, scanning staff could rescan slides with missed or out-of-focus areas, but readers could not request a rescan.

The authors “cautiously support” routine digital reporting. They make adequate reader training and quality assurance conditions for its use.

## Limits

The small numbers per specimen type prevent type-specific conclusions. The study used a single centre and a single scanner, collected no objective timing or workflow data, and used uncalibrated monitors. The team estimated the rescan rate without recording it.

For many comparisons, the glass diagnosis came from a different pathologist. The authors say using the original diagnosis in these comparisons may inflate agreement. Only cases with a digital mismatch underwent glass re-review, so the exclusion rule could favour the digital result.

Among the cases with agreement on glass re-review, the Results count six clinically significant discordant comparisons. The Discussion describes seven clinically significant discordant cases. The figures here follow the Results. The paper states that the study met the validation requirements of a US accreditation body, but it does not set out those requirements or show how each was met.

The scanner maker part-funded the study. The authors state that the maker had no role in case selection, diagnosis, analysis or drafting.

## Frequently asked questions

### Could readers ask for a fresh scan if part of a slide was out of focus?
Readers could not request a rescan during the study. Scanning staff could repeat a scan if they found missed or out-of-focus areas during their checks. A local validation can specify how readers request another scan and when they can return to glass.

### Did a deferral to glass count as a mismatch?
A deferral was left out of the reported concordance denominators. Those figures therefore describe comparisons with a digital diagnosis rather than every attempted read. Keeping the deferral count visible tells a laboratory how often its readers needed to return to glass during validation.

### What made cervical screening harder on screen?
The cervical readers found it cumbersome to work systematically across the whole digital slide. They considered screening on screen less practical in the study setup. The cervical sample was small, and the study did not measure how long screening took.