Pathologist’s Brief
Case challenge · Digital pathology

Case challenge: digital validation after equipment changes and a new organ system

The question

What should happen before the pathologist reports digitally after these equipment changes and expansion to a new organ system?

Black-and-white photo of a reporting desk with two monitors showing tissue sections, a microscope, a tray of glass slides and a signed paper form.
AI-generated illustration, Pathologist’s Brief
On this page
  1. Scenario
  2. Discussion
  3. Takeaway
  4. Source
  5. Frequently asked questions

What should happen before the pathologist reports digitally after these equipment changes and expansion to a new organ system?

Reveal the discussion

This is a fictional teaching scenario.

#Scenario

A pathologist has completed individual digital validation for a defined diagnostic scope. The laboratory introduces a new scanner and new monitors. The same pathologist now wants to start digital reporting in a new organ system.

#Discussion

The Polish Society of Pathologists guideline names both the College of American Pathologists (CAP) and Royal College of Pathologists (RCPath) validation frameworks. Its recommended approach follows the RCPath model: a pathologist-led process that identifies risks and supports safe diagnostic practice. This is one described framework, rather than the only recognised approach.

The equipment changes and the expanded scope need separate consideration. Significant technical changes, including new scanners or monitors, require revalidation scaled to the extent and nature of the change. A new organ system also requires additional validation. This follows the standard process, tailored to that diagnostic area and documented throughout.

#Archived cases

Before validation, the pathologist should become familiar with the digital system, preferably under supervision. The recommended archived set contains at least 20 anonymised cases of varied complexity, including biopsies, resections and different staining techniques. Challenging features should be represented.

Each case is assessed digitally with a complete report, then compared with the corresponding glass slides. Results and discrepancies are reviewed before progression. Additional supervised cases may be needed. Progression is agreed by the pathologist and the laboratory head or designated supervisor, ideally with an experienced instructor.

#Dual assessment in routine practice

The next stage uses routine cases. Digital assessment comes first, followed by glass-slide review before final approval. All relevant slides, including additional stains, are included. The duration and case numbers depend on experience and diagnostic scope. The guideline describes a typical period of one to three months.

Record full concordance, clinically insignificant discordance or clinically significant discordance. Document visual differences even when they do not change the diagnosis. For discordant cases, identify which method provides greater diagnostic precision and discuss the findings with the laboratory head and, where possible, the instructor.

The validation comparison is between glass slides and digital images without AI-assisted tools.

#Signed scope and outcome

A formal protocol is prepared and archived. It records the diagnostic spectrum, archived and routine case numbers, discordance analysis and participants. The outcome must state whether independent digital assessment is supported across the full spectrum, within selected areas with glass-slide use elsewhere, or not yet supported.

The laboratory head, the pathologist and the instructor, if involved, sign the protocol. Earlier validation need not be repeated when its scope still matches current practice, but it does not remove the requirements for significant technical changes or new diagnostic areas.

#Takeaway

Before the proposed reporting starts, revalidation for the new scanner and monitors and additional validation for the new organ system must be addressed. The technical revalidation is scaled to the changes, and the diagnostic scope and outcome are documented in a signed protocol.

#Source

Szylberg et al., Diagnostic Pathology, published 30 January 2026: Polish Society of Pathologists guidelines for digital pathology adoption.

Frequently asked questions

Does a new monitor mean repeating every part of validation?

The guideline scales revalidation to the extent and nature of a significant technical change. It lists new monitors and scanners among changes requiring revalidation, but does not impose the same scope for every change. The equipment update therefore needs its own assessment, alongside any additional validation for a new diagnostic area.

Does previous validation cover a new organ system?

No, expansion into a new diagnostic area requires additional validation. The guideline describes a process that mirrors standard validation, tailored to the new area and documented throughout. An earlier validation can remain relevant to an unchanged scope, but completing it does not automatically extend the pathologist to another organ system.

Can AI assistance be used during the validation comparison?

No, the recommended comparison uses glass slides and digital images without AI-assisted tools. Within this pathologist-led approach, differences between the two assessments are recorded and reviewed, including visual differences that do not alter the diagnosis. The signed protocol then records the diagnostic spectrum, discordance analysis and outcome for independent digital assessment.

Source